A Comparison of two Selective Electrodes for Vitamins B1 and B6 in terms of Fabrication and Optimal Conditions Based on Modified Carbon Paste
Keywords:
Thiamin hydrochloride, pyridoxine hydrochloride,ion selective electrodes, paste carbon electrode.Abstract
a comparative study was made between two rapid and simple electrochemical sensors from manufacturing and optimal conditions for determination the concentration of thiamine hydrochlorid (vitamin B1) and pyridoxine hydrochloride (vitamin B6). two ion exchangers was prepared as thiamine hydrochlride complex with zinc chloride and pyridoxine hydrochloride complex with cobalt chloride and used Dibutyl phthalate and paraffin oil as plasticizers.
Nernstian response of vitamin B1 selective electrode was in range concentation of (9.47×10-6-1×10-2 ) mol/L at a temperature range of (25–35)°C and a pH range of (4–6). The Nernstian slop was 28.715±2 mV/decade and a response time of 60 seconds. In contrast, the Nernstian response of vitamin B6 selective electrode was relatively narrow concentration range (1 × 10⁻⁶ - 1 × 10⁻²) mol/L and a wider temperature range (20–40°C) and lower pH range (2–4). The Nernstian slop was 56.206±2 mV/decade and a shorter response time of 45 seconds. Selectivity coefficients calculated using the matched potential method indicated that the two proposed electrodes exhibit high selectivity for vitamin B1 and vitamin B6 in the presence of various interfering ions.
The results of the two proposed electrodes were compared with two reference methods using Fisher and Student's coefficients for a rice sample. No significant differences were found between the proposed and reference methods, thus confirming the applicability of both methods for determining the concentrations of vitamins B1 and B6 .